What is the part that increases or decreases the light intensity?

When setting up a compound microscope, it is important to optimize the illumination in order to produce clear and crisp images. There are four areas that should be considered when setting up the microscope illumination.

What is the part that increases or decreases the light intensity?
#1 - Microscope Rheostat

The microscope rheostat control can be found on the side of the compound microscope body. It will typically be a knob that is turned clockwise in order to increase the light intensity, or counter-clockwise to reduce the light. The image at left shows the rheostat control on the Zeiss digital Primostar microscope circled in red. Some microscopes have a recessed dial that is rotated to the right to increase the intensity and to the left to decrease light intensity.

#2 - Using a Light Filter

Depending what type of illumination your compound microscope has, you may need to use a daylight balancing filter.  This is a blue filter and it usually sits directly on top of the illuminator, or in a filter holder above the light. If your microscope has tungsten or halogen illumination, the daylight balancing filter will correct the color temperature and result in a better quality image (and a better color image). If you have an LED microscope, this blue filter will not be needed. Microscopes that have LED illumination have a color temperature that is already high and does not need to be adjusted. The image below shows what your microscopy image may look like if you have a tungsten or halogen microscope and are not using a daylight balancing filter (left) versus what it will look like if you are using the daylight balancing filter (right).

What is the part that increases or decreases the light intensity?

#3 - Adjusting the Aperture Diaphragm

What is the part that increases or decreases the light intensity?
Beneath the microscope stage you will find the microscope condenser. The condenser has an iris diaphragm (also known as an aperture diaphragm) that can be adjusted to match the numerical aperture (NA) of the objective lens. The NA is typically typed on the side of each objective lens. Most condensers have a scale with similar numbers on them along with a lever or slider to adjust the aperture. If your objective lenses do not have the NA typed on them, below are some guidelines for adjusting the aperture diaphragm for each objective lens.

  • 4x objective = Aperture of iris should be nearly closed to 1/8 open.
  • 10x objective = Aperture of iris should be somewhere between 1/8 - 1/4 open.
  • 40x objective = Aperture of iris should be somewhere between 1/4 - 1/2 open.
  • 100x objective = Aperture of iris should be somewhere between 1/2 - 3/4 open.


#4 - Adjusting the Field Diaphragm

If your microscope has Köehler illumination, you will want to make sure you adjust the field diaphragm for optimal contrast. You can read an in-depth article here that explains how to set up your microscope for optical Köehler illumination and adjust your field diaphragm.

The four tips listed above should allow you to use your microscope to its potential - for crisp, clear, high quality images. If you have questions about adjusting your microscope illumination, contact Microscope World and we will be happy to help.

Thank you to Motic microscopes for providing the microscopy filter images.

If you’re starting your PhD or post-doctoral work, chances are you’ll need to use a light microscope at some stage during your research. Some of you may be seasoned microscopists. For many of you though, this might be the first time you’ve ever plugged in a microscope, or at least the first time you’ve used one since some long-forgotten undergraduate histology course. Yet somehow you’re expected to just dust off that old ‘scope in the corner of the lab, and start using it as if you’re an old hand. Have no fear – help is at hand! In today’s article, I’m going to go over the basics of the light microscope.

The anatomy of the light microscope

To begin with, I’d like to introduce you to the parts of the microscope – basically the bits that might be useful for you to know about. No optical physics required!

  • Microscope Slide: The glass slide that contains your specimen for examination.
  • Cover Slip: A smaller, thinner, flat piece of glass that lies over your specimen, and keeps it pressed flat against the slide to maintain an even thickness. This makes it easier to focus on the specimen.
  • Stage: The platform where you place the microscope slide. It contains a slide holder, which is a spring-loaded clip that holds your slide firmly in place on the stage.
  • Stage Controls: Dials for controlling the position of the stage. These allow you to move the slide around to bring different regions of the image into view.
  • Coarse focus: This dial is used to bring the image roughly into focus.
  • Fine Focus: This dial is then used to more finely tune the image into clear focus.
  • Eyepieces: Also called the oculars, or ocular lenses. This is the set of lenses at the top that you look through. They collect light and focus it to produce the image that you see. They also take the image produced by the objective lens and magnify it another ten times.
  • Nosepiece: A rotating carousel above the stage that contains the objective lenses.
  • Objective lenses: Three or four  objective lenses of different magnifications will be attached to the nosepiece – you can flip between these to view your image at different magnifications.
  • Iris Diaphragm: A rotating disc under the stage that controls the intensity of light hitting the specimen.
  • Condenser: Usually sits just above the iris diaphragm. Focuses light onto the specimen.
  • Illuminator: The light source, which illuminates the specimen.
  • Light Control Dial: Controls the brightness of light heading toward the specimen.

What is the part that increases or decreases the light intensity?

Viewing your slides

OK, so now that you know all the parts of the microscope, let’s look at your first slide. What do you do first?

  • Switch on the microscope: I know it seems intuitive, but trust me, I’ve seen many people become frustrated at not being able to see anything, only to discover they hadn’t plugged it in or switched it to the “on” position.
  • Set the slide on the stage: Allow the mechanical spring-loaded clip to grasp it in place. Make sure it’s the right way up – the cover slip should be facing upwards.
  • Look through the eyepieces: Adjust the width between the eyepiece lenses to match that of your eyes (I’m keeping my fingers crossed that they’ve given you a binocular microscope, and not a monocular one!). Use the light control dial to set a comfortable light intensity. This will vary from person to person – aim for a brightly illuminated image, but not so bright that your retinas feel like they’re on fire!
  • Start at the lowest magnification: Rotate the nosepiece to click the lowest magnification objective lens into place – this is the shortest lens (yours may be 2x or 4x). Use the coarse focus dial to lower the lens as close to the slide as possible, without touching it. Take care to look at the slide and lens as you do this – it’s easy to crush the slide. Once the slide and lens are close together, look through the eyepieces again to focus. Use the coarse focus dial first – turn it slowly in the opposite direction than you previously did (i.e. away from the slide) until you can see a clear image. Then use the fine focus dial to bring your image into sharp focus.
  • Examine your specimen at progressively increasing magnifications: Once your image is in focus at low magnification, you can switch objective lenses and your image will remain mostly in focus – only fine focus adjustments may be required. Move up through the objective lenses in turn, to evaluate your specimen from lowest to highest magnification.

Hopefully, this overview has taken some of the mystery out of using your lab’s ancient light microscope. It can be a nightmare if you’re faced with using a microscope for the first time, but following these few, easy steps should ease the pain a little!

What part of the microscope increases or decreases light intensity?

Condenser Diaphragm- This diaphragm controls the amount of light entering the lens system. This feature is useful for viewing unstained biological specimens that are translucent. Reducing the amount of light improves contrast, making the specimen "stand out" against the background.

What part of the microscope controls the light intensity?

Iris Diaphragm: A rotating disc under the stage that controls the intensity of light hitting the specimen.

Does the light intensity increase or decrease as the magnification increases?

When an individual increases the magnification of objects viewed on the microscope stage, the light intensity gradually reduces. This occurs because magnification focuses on a smaller area, and hence the light gets scattered and specimen becomes less focused.

Which part of the microscope is used to adjust the intensity brightness of the light passing through the specimen on the stage?

Light from the light source then passes through the condenser which focuses the light on the specimen. An iris diaphragm is used to control the intensity, or brightness, of light which passes through the specimen, thus allowing the operator to adjust the intensity and achieve an optimum viewing contrast.