The objective (lens) is the first set of optical systems that image the object being observed, and is also the most important imaging component in the microscope. Biological Objective Show Some objectives are mounted directly on the microscope body, some separate from the body and are installed when needed. Usually, on the objective outer casing, there are signs of the following parameters: Usually, the objective has very fine mounting threads. When there is a need to install the objective /objective frame, be careful to install it. Align the nosepiece installation position, keep it completely “flat”. When it is blocked, remove it and reinstall it. Do not force it in. Note: although between different manufacturers, some objectives can be used universally, they may still bring magnification error and image quality degradation. Microscope MagnificationA microscope's total magnification is a combination of the eyepieces and the objective lens. For example, a biological microscope with 10x eyepieces and a 40x objective has 400x magnification. There are however, a few limits to the amount of total magnification that can be reached before empty magnification comes into play. Empty magnification occurs when the image continues to be enlarged, but no additional detail is resolved. This is often the case when higher magnification eyepieces are used. In order to avoid empty magnification, there are a few simple steps that are helpful to follow. Eyepiece and Objective Combinations for Optimal MagnificationWhen selecting a combination of eyepieces and objective lenses for the optimal magnification, without ending up with "empty magnification" it is important to consider the numerical aperature (NA) of the objective. The numerical aperture of a microscope objective defines the objective's resolution. Each microscope objective has a minimum and maximum magnification necessary for the details in an image to be resolved. A simple formula for the minimum value is (500 x NA). And for the maximum magnification (1000 x NA). Magnifications higher than this value will result in empty magnification, or an image that has a poor resolution. The table below shows some typical NA values with their corresponding objective and provides a range of useful magnification combinations. The blank boxes in the table would provide empty magnification and should be avoided. For example, pairing 20x eyepieces with a 100x objective would not provide good resolution and would result in empty magnification. To determine this, we took 1.25NA x 1000 = 1250 magnification maximum. However, the combination of the 100x objective x 20x eyepieces = 2000, which is above the maximum magnification. Range of Useful Magnification based on NA of Objectives
If you have any questions about your microscope's magnification, contact Microscope World. What does 40x mean on a microscope?A 40x objective makes things appear 40 times larger than they actually are. Comparing objective magnification is relative—a 40x objective makes things twice as big as a 20x objective while a 60x objective makes them six times larger than a 10x objective. The eyepiece in a typical desktop microscope is 10x.
What is a 4x lens known as?Scanning Objective Lens (4x) Low Power Lens (10x) High Power Lens (40x) Oil Immersion Lens (100x)
Why is the 40x objective sometimes called the high dry objective?The 40x objective is known as a high dry lens.
This is a high magnification, hence the "high" part of high dry. However, unlike the higher 100x objective lens, the 40x lens is not used with oil immersion microscopy, and therefore is "dry."
What is the magnification of 40x on a microscope?Field of view is how much of your specimen or object you will be able to see through the microscope. At 40x magnification you will be able to see 5mm. At 100x magnification you will be able to see 2mm. At 400x magnification you will be able to see 0.45mm, or 450 microns.
|